TRACE: Two-Stage Detector-Response Estimation With Angular Cosine Expansion for Ring Artifact Correction in Photon-Counting CT
Paper recorded by Signals 4 on 2026-09-14 in cs.CV. Abstract reproduced from arXiv; link to the original below.
Published 2026-09-14 on arXiv · recorded by Signals 4 on 2026-09-15
Category: cs.CV · 计算机视觉 · first seen 2026-09-15
Abstract
Detector response nonuniformity introduces systematic projection errors and ring artifacts in photon-counting detector computed tomography (PCD-CT). In measured PCD-CT data, residual stripe amplitudes vary slowly with projection angle, which fixed-bias models cannot adequately capture. We propose TRACE, a two-stage unsupervised sinogram decomposition method for estimating and correcting these resp
Read on arXiv →
Cite this page: TRACE: Two-Stage Detector-Response Estimation With Angular Cosine Expansion for Ring Artifact Correction in Photon-Counting CT: the #46 most recent of 237 cs.CV papers we have recorded (as of 2026-09-14). Source: Signals 4 (Signals API) — https://data.jiangzhang.ca/signals4/t/papers/trace-two-stage-detector-response-estimation-with-angular-cosine-expansion-for-r.html